ResMed AirSense 11 Error Codes & Messages

Browse the verified ResMed faults that explicitly identify AirSense 11 in their model context. A model hub is shown only when the record set contains useful coverage.

22 verified faults

Faults for AirSense 11

  • ResMed High Leak Detected display message
    High Leak Detected means the device is sensing more unintended air loss than expected. The manufacturer guidance points to mask fit, tubing connections, and—on …
  • ResMed Error 4 numbered error
    For AirSense 11, ResMed's official user guide links Error 4 to a hot environment, a blocked air filter, or water in the air tubing. The associated service guide…
  • ResMed Tubing Blocked display message
    The device is detecting or reporting restricted airflow through the external air tubing. ResMed's user guidance directs the user to check the tubing and remove …
  • ResMed Humidifier Fault display message
    Humidifier Fault identifies a problem in the humidifier system, but the user consequence depends on the device family. The AirSense 11 guide says therapy can ru…
  • ResMed Error 9 numbered error
    Motor H bridge failure; faulty main PCB - replace main PCB…
  • ResMed Error 8 numbered error
    Motor sticky; faulty motor - replace motor…
  • ResMed Error 7 numbered error
    Motor hardware out of bounds; faulty motor - replace motor…
  • ResMed Error 6 numbered error
    Motor stall (software, back EMF check); faulty motor or faulty main PCB - replace faulty part…
  • ResMed Error 5 numbered error
    Motor over-voltage; faulty power supply, faulty motor or faulty main PCB - replace faulty part…
  • ResMed Error 3 numbered error
    Mask over pressure (severe) / Motor fast over pressure; pressure sensor fault or software issue - complete pressure test, replace main PCB…
  • ResMed Error 24 numbered error
    Faulty HW fault detection circuitry; faulty main PCB - replace main PCB…
  • ResMed Error 23 numbered error
    Implausible supply voltage; pressure sensor issue - replace main PCB…
  • ResMed Error 21 numbered error
    Mixed signal chip error; faulty main PCB - replace main PCB…
  • ResMed Error 18 numbered error
    Calibration reset; data corruption - power cycle device, test device and recalibrate, send device logs to ResMed for analysis…
  • ResMed Error 17 numbered error
    Settings reset; flash corruption or file system error - power cycle device, send device logs to ResMed for analysis…
  • ResMed Error 15 numbered error
    Pressure sensor stuck low; pressure sensor issue - complete device testing in RSC, replace main PCB…
  • ResMed Error 14 numbered error
    Pressure sensor stuck high; pressure sensor issue - complete device testing in RSC, replace main PCB…
  • ResMed Error 13 numbered error
    Pressure sensor drift 5cm; pressure sensor issue - complete device testing in RSC, replace main PCB…
  • ResMed Error 12 numbered error
    No flow data; faulty main PCB - replace main PCB…
  • ResMed Error 11 numbered error
    Flow sensor stuck high; faulty main PCB - replace main PCB…
  • ResMed Error 10 numbered error
    Flow sensor stuck low; faulty main PCB - replace main PCB…
  • ResMed Error 1 numbered error
    Motor has stalled (hardware and MT4 Over Temperature); faulty motor - replace motor…

Model-matched external accessories

These are user-accessible accessories, not repairs. Confirm the exact device label and accessory type before ordering.

  • AirSense 11-compatible external air filters — External user-replaceable air filters only. Confirm AirSense 11 compatibility; this does not repair an internal device fault. View on Amazon →
  • ResMed AirSense 11-compatible replacement tubing — External tubing only. Confirm AirSense 11 connector and exact tube type; a replacement tube does not repair an internal fault. View on Amazon →
  • AirSense 11-compatible HumidAir replacement water chamber — External humidifier tub only. Confirm AirSense 11 compatibility and tub type; it does not repair an internal humidifier fault. View on Amazon →